Automated screening of blood cultures with the Malthus microbiological growth analyser.
Clinical Microbiology and Public Health Laboratory, Addenbrooke's Hospital, Cambridge.
A total of 3347 blood cultures from patients in all hospital wards were examined on a Malthus microbiological growth analyser and by a conventional system. There was no significant difference in the total numbers of positive cultures of clinical importance between the two systems (p greater than 0.05). Staphylococcus aureus, however, was isolated more often by the conventional method (p less than 0.05). Failure of the automatic detection routine limited the potential of the Malthus system for earlier detection of positive cultures. Daily visual examination of Malthus curves and subculture of bottles not promptly attached to the apparatus were necessary to avoid missing some positive cultures. False positive rates were 13% for the Malthus system and 2% for the conventional system. The contamination rate was considerably lower in the Malthus system (p less than 0.001). Further development would be necessary for the apparatus to be acceptable for routine screening of blood cultures.
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.