Comparison of conventional and single bottle system for blood cultures.
Department of Microbiology, Central Pathology Laboratory, Stoke-on-Trent.
A single bottle blood culture system (Oxoid Signal system) was compared with a conventional two bottle subculturing system. A total of 2016 routine blood samples yielded 186 (9%) clinically important isolates. Of these, 40 (21%) were isolated only in the Oxoid system and 30 (17%) only in the conventional system. One hundred and sixteen (62%) were isolated from both systems. Volume of blood was not significantly associated with the rate of detection but was significantly associated with the speed of detection. A continuation of the study with a Signal bottle and single conventional bottle confirmed initial results of the increased isolation rate using two systems. The Oxoid single bottle blood culture system was at least comparable with a conventional system in terms of rate of detection of organisms and was superior in terms of speed of detection.
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.