Evaluation of sealed vacuum extraction method (Seditainer) for measurement of erythrocyte sedimentation rate.
Department of Haematology, Medical School, University of Birmingham.
A sealed vacuum extraction method (Seditainer) for determining the erythrocyte sedimentation rate (ESR) was compared with the standard Westergren ESR technique. The Seditainer method was particularly easy to use, showed acceptable precision, reduced the biohazard risk to laboratory staff and, on storage of sealed blood specimens for 24 hours at 4 degrees C, gave ESR values that had decreased by a mean of only 1.9% (95% CI + 0.2 to -4.0%). Seditainer tubes are shorter (100 mm) than Westergren tubes (200 mm) and this reduced test sensitivity at ESR values above 55 mm/first hour. After application of the manufacturer's correction formula to adjust for tube length there was improved correlation (n = 150, r = 0.936, p less than 0.001), but still considerable scatter, between the Seditainer and Westergren results. Seditainer ESR values above 55 mm/first hour should therefore be regarded as semiquantitative. This vacuum extraction method offers a simple and safe technique for measuring the ESR and specimens can be stored overnight at 4 degrees C.
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.