© 2000 Journal of Clinical Pathology
Short report
Clinical use of WHO haemoglobin colour scale: validation and critique
1 Helen Joseph Hospital Laboratory, South African Institute for Medical Research, PO Box 1038, Johannesburg 2000, South Africa
2 WHO Collaborating Centre for Haematology Technology, Department of Haematology, Hammersmith Hospital, Imperial College School of Medicine, London W12 0NN, UK
Correspondence to:
Dr Lewis smlenl{at}intonet.co.uk
AimThe World Health Organisation (WHO) haemoglobin colour scale has been developed as a simple, inexpensive clinical device for diagnosing anaemia when laboratory based haemoglobinometry is not available. In an initial validation study at several health centres, scale readings were compared with measurements of haemoglobin by the laboratory. This showed the scale to have 90% sensitivity and 70% specificity in identifying whether anaemia was present or not. In addition, when present, the degree of anaemia was correctly classified in clinical terms as moderate, pronounced, or severe, with an overall sensitivity of 60% and specificity of 88%. Errors were mainly marginalthat is, between two adjacent categoriesbut there were also some major discrepancies, such as a blood with a haemoglobin of 67 g/dl being read as normal or vice versa. Because this would compromise the scale's reliability in practice, this study was undertaken to identify the causes of the discrepancies and to reassess the performance of the scale.
MethodsVenous blood samples were collected into potassium EDTA from patients attending selected clinics at three South African hospitals with good laboratory facilities. A prototype of the device was used unsupervised by nursing staff, doctors, and phlebotomists, who were told to follow the printed instructions. The blood specimens were then immediately sent to the laboratory where haemoglobin was measured by standardised automated blood cell counters. Any discrepancies > 1 g/dl were recorded and the tests were repeated by the same operators under supervision of the investigators.
ResultsAlmost all the errors that occurred resulted from the incorrect use of the device, namely: inadequate or excessive blood, reading the results too soon or too late (beyond the limit of two minutes), poor lighting, or holding the scale at the wrong angle. The accuracy improved dramatically when the tests were repeated under supervision and these faults were avoided: 95% of readings were within 1 g/dl of the reference measurements, and 97% within 1.5 g/dl. Anaemia screening showed 96% sensitivity and 86% specificity. Clinical judgement of pallor was frequently wrong, whereas the scale gave the correct diagnosis in more than 97% of cases.
ConclusionThe study confirmed the usefulness and reliability of the scale and its advantage over clinical signs for the diagnosis of anaemia, thus providing a clinically reliable near patient method in the absence of a laboratory. The instructions are easy to follow but must be strictly adhered to.
Key Words: diagnosis of anaemia haemoglobinometry haemoglobin colour scale clinical assessment of anaemia
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
-
Critchley, J., Bates, I.
(2005). Haemoglobin colour scale for anaemia diagnosis where there is no laboratory: a systematic review. Int J Epidemiol
34: 1425-1434
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.