Register for email alerts and news feeds:
This journal | BMJ Group
rss
Journal of Clinical Pathology 2004;57:378-382; doi:10.1136/jcp.2003.012294
Copyright © 2004 by the BMJ Publishing Group Ltd & Association of Clinical Pathologists.
Journal of Clinical Pathology 2004;57:378-382
© 2004 BMJ Publishing Group Ltd & Association of Clinical Pathologists

ORIGINAL ARTICLE

Analytical and diagnostic accuracy of "second generation" assays for thyrotrophin receptor antibodies with radioactive and chemiluminescent tracers

D Villalta1, E Orunesu2, R Tozzoli3, P Montagna2, G Pesce2, N Bizzaro4, M Bagnasco2

1 Clinical Immunology and Virology Unit, City Hospital Pordenone, Via Montereale 24, 33170 Pordenone, Italy
2 Autoimmunity Laboratory, Medical and Radiometabolic Therapy Unit, DI.M.I. University of Genoa, Viale Benedetto XV 6, 16132 Genoa, Italy
3 Clinical Pathology Unit, City Hospital Latisana, Via Sabbionera 45, 33053 Latisana, Italy
4 Clinical Pathology Unit, City Hospital San Donà di Piave, Via N. Sauro 25, 30027 San Donà di Piave, Italy

Correspondence to:
Correspondence to:
Dr M Bagnasco
Autoimmunity Laboratory, Medical and Radiometabolic Therapy Unit, DI.M.I. University of Genoa, Viale Benedetto XV 6, 16132 Genoa, Italy; allerlab{at}unige.it

Aims: To investigate the analytical and diagnostic accuracy of thyrotrophin (TSH) receptor antibody assays using recombinant human TSH receptors.

Methods: Sera from 68 patients with Graves’ disease, 23 patients with autoimmune thyroiditis, and 119 healthy controls were evaluated in four different laboratories using both radioactive and chemiluminescent tracers. Functional sensitivity, interlaboratory precision, optimal cutoff values for Graves’ disease, and the correlation between the two methods were evaluated.

Results: Functional sensitivity was 0.98 IU/litre for both assays. Interlaboratory precision, expressed as per cent coefficient of variation over a wide range of antibody concentrations, varied from 5.7% to 15.1% for the radioligand, and from 6.6% to 19.9% for the chemiluminescence assay. The two methods (radioactive and chemiluminescent) were closely correlated. All the sera from untreated or relapsing patients with Graves’ disease gave TSH receptor antibody values above 2.1 IU/litre, whereas in none of the healthy controls did values exceed 2.5 IU/litre. Receiver operating curve analysis allowed an optimal cutoff point to be defined at 1.99 IU/litre, according to a sensitivity of 100% and specificity of 99.1%.

Conclusions: These data show the high analytical and diagnostic accuracy of the human TSH receptor assays, both with radioactive and chemiluminescent tracers, when both functional sensitivity and interlaboratory reproducibility are considered. These two methods could be proposed as first line diagnostic markers for Graves’ disease.

Abbreviations: CI, confidence interval; CV, coefficient of variation; GD, Graves’ disease; TBII, thyrotrophin binding inhibitory immunoglobulins; TRAb, autoantibodies to the thyrotrophin receptor; TSH, thyrotrophin


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

This article has been cited by other articles:

  • Jensen, E., Petersen, P. H., Blaabjerg, O., Hansen, P. S., Hegedus, L. (2005). Improved Sensitivity of a Thyrotropin Receptor Antibody Assay. Clin. Chem. 51: 2186-2187 [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Pathology jobs

Pathology jobs