Comparing substrates for the detection of ANAs
1 Central Sydney Immunology Laboratory, Royal Prince Alfred Hospital, Missenden Road, Camperdown, NSW 2050, Australia
The article by Pollock and Toh1 about the detection of antinuclear antibodies (ANAs) using HEp-2 cells transfected with Ro/SS-A raises the important question of the optimal method for screening of sera from patients with suspected autoimmune connective tissue disease. We have directly compared the performance of the same Hep-2 transfected cells (Hep2000; Immuno Concepts, California, USA) with Hep-2 untransfected cells (Quantafluor; Sanofi Diagnostics Pasteur Inc, Minnesota, USA). The results from our study combined with a reassessment of Pollock and Toh's data cast doubt on their conclusion that Hep-2 transfected cells are more reliable than other substrates for detecting clinically meaningful ANAs.
Sera from 258 patients referred to our laboratory for ANA testing were analysed for the presence, titre, and pattern of ANAs using Ro/SS-A transfected and untransfected HEp-2 cells. Indirect immunofluorescence was performed at a screening dilution of 1/40 and titred to 1/2560, and slides were reviewed independently by at
2 Department of Pathology and Immunology, Monash Medical School, Alfred Hospital Campus, Commercial Road, Prahran, Victoria 3181, Australia
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.