Article Text

Download PDFPDF
An evaluation of the Vickers Instruments J12 cell counter
  1. D. E. Pegg,
  2. A. C. Antcliff1
  1. Westminster Medical School and the Westminster Hospital, London

    Abstract

    The Vickers J12 cell counter uses an optical hydraulic counting system to perform rapid automatic counts of leucocytes and erythrocytes in liquid suspension. The effect of sample storage, the use of different diluents, variation in cell dimensions and other relevant factors have been studied. A counting accuracy of ± 1·19% for replicate erythrocyte counts and ± 2·94% for leucocyte counts was obtained.

    Statistics from Altmetric.com

    Request Permissions

    If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

    Footnotes

    • 1 Present address: Department of Chemical Pathology, Postgraduate Medical School, Hammersmith, London.