A powered bone marrow biopsy system versus manual methods: a systematic review and meta-analysis of randomised trials
  1. Jeffrey Voigt1,
  2. Michael Mosier2
  1. 1Medical Device Consultants of Ridgewood, LLC, Ridgewood, New Jersey, USA
  2. 2Department of Mathematics and Statistics, Washburn University, Topeka, Kansas, USA
  1. Correspondence to Jeffrey Voigt, Medical Device Consultants of Ridgewood, LLC, 99 Glenwood Rd., Ridgewood, NJ 07450, USA; meddevconsultant{at}aol.com

Abstract

A significant number of bone marrow biopsies are performed annually. Outcomes of patient pain and sample size of biopsy are important issues as perception of patient pain is underestimated by clinicians and sample size assists in accurate diagnosis. Manual extraction of marrow biopsy is the main method used. Recently a powered system has been introduced. The objectives of this systematic review and meta-analysis were to determine if the powered system reduces patient pain and improves sample capture. A PubMed and Cochrane search for randomised controlled trials was undertaken comparing the powered system with manual methods. Risk of bias assessment and meta-analysis of outcomes using appropriate statistical methods were performed. Five randomised controlled trials were identified. Patient pain (measured via visual analogue scale (VAS)—100 point scale) was significantly reduced using the powered system: mean difference=−6.57; 95% CI −12.93 to −0.22; p=0.04. The relative reduction in pain was 17%–25% with the powered system. Sample biopsy size (length in mm) was also significantly increased with the powered system: mean difference=3.65 mm; 95%; CI 1.61 mm to 5.68 mm; p=0.0005. The relative increase in sample size was 33% with the powered system. Operator ease of use (as measured via VAS) and adverse events were similar. Despite limited operator experience, patients experienced less pain and sample sizes were increased without an increase in adverse events with the powered system. The powered system may offer an option in obtaining samples from patients whose pain is compromised by disease and may reduce redos.

  • Bone Marrow Trephines
  • Bone Tumour Pathology
  • Myeloma

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 3.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/3.0/

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Supplementary materials

  • Supplementary Data

    This web only file has been produced by the BMJ Publishing Group from an electronic file supplied by the author(s) and has not been edited for content.

    Files in this Data Supplement:

Linked Articles

  • Correction
    BMJ Publishing Group Ltd and Association of Clinical Pathologists