Table 1

Results of the detection of Chlamydia trachomatis in male first void urine and female endocervical swab specimens by IDEIA, IDEIA PCE, and Amplicor PCR

ProceduresFinal results
Sample sourceNo. of specimensIDEIAIDEIA PCEAmplicor PCRPositiveNegative
Total no. refers to the number of confirmed positive cases by each assay.
*All five were positive by confirmatory nested PCR.
†Of three specimens, two were positive and one negative by confirmatory nested PCR.
IDEIA, enzyme immunoassay; PCE, polymer conjugate enhanced; PCR, polymerase chain reaction.
Male first void urine1350135
46+++460
7++70
5*+50
Total no.46535858135
(n = 193)
Female endocervical swab1460146
34+++340
4++40
3†+21
Total no.34384140147
(n = 187)