Table 2

Mean international normalised ratio (INR) and interlaboratory variation (CV) in two separate external quality assessment surveys. INR values were calculated and reported by each participating laboratory

Sample A not illuminatedSample B not illuminatedSample C illuminated
ReagentnMeanCV (%)MeanCV (%)MeanCV (%)
The artificial blood samples were prepared separately for each survey. Artificial blood sample A was prepared from pooled coumarin plasma without addition of methylene blue and without illumination. Blood sample B was prepared from the same pooled coumarin plasma but with addition of methylene blue and without illumination. Blood sample C was prepared from the same pooled coumarin plasma but with the addition of methylene blue and with illumination.
n, number of laboratories.
Thrombotest413.114.63.114.93.134.5
Thrombotest402.874.32.894.32.934.3
Recombiplastin183.558.53.528.03.758.4
Recombiplastin193.348.03.387.03.717.5
PT-Fib HS173.7312.03.7911.94.2413.9
PT-Fib HS193.5712.03.6410.94.1614.0
Thromborel-S133.704.63.695.63.925.7
Thromborel-S123.566.83.627.13.907.6
Hepato Quick163.105.73.106.33.357.1
Hepato Quick163.125.63.155.23.365.1