Table 1

Characteristics of the commercial (A–H) and in house assays for PR3-ANCA

KitUnitsNumber of standardsDynamic rangeSample dilutionData reductionAssay performance
SampleIntra-assay CV%Interassay CV%
CV, coefficient of variation; I/H, in house; OD, optical density; PR3-ANCA, proteinase 3 antineutrophil cytoplasmic antibodies.
AUnitsLow pos1/101(OD samples/OD low positive) ×ELISA low positiveNegative34
Low29
High32
BU/ml60–1001/100Log/linear 4 parameter14.76.2
22.88.8
32.83.9
CU/ml70–1001/51Best fit curve16.37.3
26.89.6
DUnits524–3051/40Linear best fit curve
EU/ml60–1001/101Linear plot of normalised OD: OD mean/OD max standard %113.37.6
212.15.6
313.95.3
FEU/ml421–1701/101Linear/linear best fit curve110.68.4
26.58.7
GEU410–3001/50Linear/linearLow1014.7
Medium2.18
High12.411.9
Negative00
HAU/ml55–1001/50Log/linearLow8.19
Medium4.75.2
High2.65.7
I/HU/ml80–1001/50Log/linear 4 parameterLow57
High2124