The Reiter protein complement-fixation test using the AutoAnalyzer
  1. V. W. Pugh,
  2. R. W. T. Gaze
  1. Department of Clinical Pathology, The Royal Infirmary, Leicester

    Abstract

    A method for the Reiter protein complement-fixation test using the Techncon AutoAnalyzer is described. Two hundred and fifty sera were tested, when six false-positive and nine false-negative results were obtained when compared with results obtained by the manual method. Deep freezing the sera before testing on the AutoAnalyzer increased the sensitivity so that in a further 250 sera tested no false-negative and two false-positive results were obtained when compared with the results obtained at the Venereal Diseases Reference Laboratory, London.