Article Text
Abstract
Background: The need for urgent antineutrophil cytoplasmic antibody (ANCA) results when assessing patients with acute renal failure, pulmonary renal syndrome, or mononeuritis multiplex has led to the development of a rapid qualitative ELISA screening assay for antibodies to myeloperoxidase (MPO) and proteinase 3 (PR3).
Aims: To report the use of a rapid qualitative ELISA screen for PR3-ANCA and MPO-ANCA in a regional immunology laboratory and its correlation with standard indirect immunofluorescence (IIF) and quantitative ELISA for PR3-ANCA and MPO-ANCA.
Methods: Over 12 months, 103 samples requiring urgent ANCA testing were screened by a rapid qualitative ELISA and the results compared with IIF and quantitative ELISA assays for PR3-ANCA and MPO-ANCA.
Results: There was an excellent correlation between the rapid qualitative ELISA and standard ANCA IIF and a routine ELISA for MPO/PR3-ANCA, with sensitivities ranging from 82% to 100%. There were two false negatives, which gave weak to moderately positive values as determined by routine ELISA. However, the clinical relevance of these two cases is doubtful.
Conclusions: The rapid ELISA for anti-MPO and anti-PR3 correlates well with quantitative ELISA and IIF ANCA, and urgent management decisions in patients with suspected small vessel vasculitis can be based with confidence on this test.
- antineutrophil cytoplasmic antibodies
- antimyeloperoxidase antibodies
- antiproteinase 3 antibodies
- vasculitis
- enzyme linked immunosorbent assay
- ANCA, antineutrophil cytoplasmic antibodies
- ELISA, enzyme linked immunosorbent assay
- IIF, indirect immunofluorescence
- MPO, myeloperoxidase
- OD, optical density
- PR3, proteinase 3