TY - JOUR T1 - An evaluation of the Lancer Coagulyzer in the measurement of the one-stage prothrombin time. JF - Journal of Clinical Pathology JO - J Clin Pathol SP - 231 LP - 236 DO - 10.1136/jcp.29.3.231 VL - 29 IS - 3 AU - L M Rowlands AU - A C Lawrence Y1 - 1976/03/01 UR - http://jcp.bmj.com/content/29/3/231.abstract N2 - An evaluation of a photoelectric clot timer, the Lancer Coagulyzer, in measuring the one-stage prothrombin time has been carried out. The machine is considered to provide reliable results on correctly taken samples without interference by plasma bilirubin, only rarely failing to detect clotting in turbid plasma using a wide range of thromboplastins. The apparatus is useful where there is a sufficient number of samples to be tested. ER -