TY - JOUR T1 - Method for the measurement of antioxidant activity in human fluids JF - Journal of Clinical Pathology JO - J Clin Pathol SP - 356 LP - 361 DO - 10.1136/jcp.54.5.356 VL - 54 IS - 5 AU - D Koracevic AU - G Koracevic AU - V Djordjevic AU - S Andrejevic AU - V Cosic Y1 - 2001/05/01 UR - http://jcp.bmj.com/content/54/5/356.abstract N2 - Aim—To develop a new, simple, and cheap method for estimating antioxidant activity in human fluids. Methods—The assay measured the capacity of the biological fluids to inhibit the production of thiobarbituric acid reactive substances (TBARS) from sodium benzoate under the influence of the free oxygen radicals derived from Fenton's reaction. A solution of 1 mmol/litre uric acid was used as standard. Results—The following mean (SD) antioxidative activities were found (as uric acid) in the various biological fluids: serum, 2.04 (0.20) mmol/litre; urine, 176.5 (25.6) μmol/litre; cerebrospinal fluid, 95.0 (26.9) μmol/litre; aqueous humour oculi, 61.25 (9.9) μmol/litre; saliva, 838.5 (48.2) μmol/litre; tears, 247.0 (17.0) μmol/litre; ascites fluid, 270.0 (63.3) μmol/litre; kidney cyst fluid, 387.1 (28.1) μmol/litre. Small samples of the biological material were needed for the analyses: 10 μl of serum and 50–100 μl of other body fluids. In the sera of 48 healthy individuals there was a significant positive correlation between values obtained with the Randox method (as a reference method) and the new method proposed here (correlation coefficient, 0.8728; mean difference between methods, <0.4%). Conclusions—This method is easy, rapid, reliable, and practical for the routine measurement of total antioxidant activity in serum and other human body fluids. Small samples of biological material are needed for the analyses and the results are comparable with the reference (Randox) method. ER -