Results of the κ reliability test comparing immunofluorescence (IFL), enzyme linked immunosorbent asssay (ELISA), Varelisa, LKM1-ELISA, and radioligand assay (RLA) readings
IFL | ELISA | Varelisa | LKM1- ELISA | RLA | |
---|---|---|---|---|---|
A κ value of 0.8 to 1.0 denotes very good agreement, 0.6 to 0.8 substantial agreement, and 0.4 to 0.6 moderate agreement. | |||||
IFL | 1.0 | ||||
ELISA | 0.98 | 1.0 | |||
Varelisa | 0.95 | 0.92 | 1.0 | ||
LKM1-ELISA | 0.67 | 0.65 | 0.68 | 1.0 | |
RLA | 0.98 | 1.0 | 0.92 | 0.67 | 1.0 |