Table 1

 Accuracy of various haemoglobin methods compared with the reference haemiglobincyanide (HCN) method

MethodMean (g/l)Median (g/l)SDRange95% limit below reference method95% limit above reference method
One hundred different samples were used for each method.
HemoCue94104461.6–16.56%16%
HCN94106461.9–16.9
Jenway colorimeter6559301.9–16.114%10%
HCN6760302.1 –16.3
WBA colorimeter5852251.9–14.531%31%
HCN6156281.9–16.0
Haemoglobin colour scale8180354.0–14.048%37%
HCN7872352.3–17.0
DHT meter8177302.8–16.819%33%
HCN7568301.9–15.4
Lovibond105110364.0–14.083%33%
HCN8178401.5–16.5