Table 2

 Discrepancies seen when scoring the HercepTest by eye or using the MDS system

SampleHercepTest scoreFISH resultComments
By eyeMDS
FISH, fluorescence in situ hybridisation; MDS, Micrometastasis Detection System.
a23AmplifiedWould analyse by FISH routinely but MDS accurate
b21Non-amplifiedMDS accurate
c01Amplified2 tumour populations seen: 1 amplified
d32Non-amplifiedMDS accurate
e23AmplifiedWould analyse by FISH routinely but MDS accurate
f21Non-amplifiedMDS accurate
g21Non-amplifiedMDS accurate
h21Non-amplifiedMDS accurate
i33Non-amplifiedHercepTest staining inaccurate
j33Non-amplifiedHercepTest staining inaccurate
k21Non-amplifiedMDS accurate
l21Non-amplifiedMDS accurate
m21Non-amplifiedMDS accurate
n21AmplifiedMDS does not correspond to FISH
o12Non-amplifiedMDS accurate
p21Non-amplifiedMDS accurate
q23AmplifiedWould analyse by FISH routinely but MDS accurate
r23AmplifiedWould analyse by FISH routinely but MDS accurate
s12AmplifiedMDS accurate. Would not analyse by FISH routinely
t21Non-amplifiedMDS accurate
u13Amplified2 tumour populations seen: 1 amplified