Table 3

Percentage agreement and 95% CIs in the interlaboratory reproducibility study

EvaluationPercentage agreement (95% CI)
Intraobserver reproducibilityANA: 92.1 (89.6 to 94.4)
APA: 94.2 (92.4 to 95.9)
OPA: 93.3 (91.3 to 95.2)
Interobserver reproducibilityANA: 90.2 (88.7 to 91.7)
APA: 92.9 (91.7 to 94.0)
OPA: 91.8 (90.5 to 93.0)
Intralaboratory reproducibilityANA: 95.1 (93.3 to 96.7)
APA: 96.0 (94.5 to 97.3)
OPA: 95.6 (94.0 to 97.1)
Interlaboratory reproducibilityANA: 93.2 (91.9 to 94.4)
APA: 94.4 (93.4 to 95.5)
OPA: 93.9 (92.7 to 94.9)
  • ANA, average negative agreement; APA, average positive agreement; OPA, overall percentage agreement.