Efficacy of implantable cardioverter-defibrillators for the prevention of sudden death in patients with hypertrophic cardiomyopathy

N Engl J Med. 2000 Feb 10;342(6):365-73. doi: 10.1056/NEJM200002103420601.

Abstract

Background: Hypertrophic cardiomyopathy is a genetic disease associated with a risk of ventricular tachyarrhythmias and sudden death, especially in young patients.

Methods: We conducted a retrospective multicenter study of the efficacy of implantable cardioverter-defibrillators in preventing sudden death in 128 patients with hypertrophic cardiomyopathy who were judged to be at high risk for sudden death.

Results: At the time of the implantation of the defibrillator, the patients were 8 to 82 years old (mean [+/-SD], 40+/-16), and 69 patients (54 percent) were less than 41 years old. The average follow-up period was 3.1 years. Defibrillators were activated appropriately in 29 patients (23 percent), by providing defibrillation shocks or antitachycardia pacing, with the restoration of sinus rhythm; the average age at the time of the intervention was 41 years. The rate of appropriate defibrillator discharge was 7 percent per year. A total of 32 patients (25 percent) had episodes of inappropriate discharges. In the group of 43 patients who received defibrillators for secondary prevention (after cardiac arrest or sustained ventricular tachycardia), the devices were activated appropriately in 19 patients (11 percent per year). Of 85 patients who had prophylactic implants because of risk factors (i.e., for primary prevention), 10 had appropriate interventions (5 percent per year). The interval between implantation and the first appropriate discharge was highly variable but was substantially prolonged (four to nine years) in six patients. In all 21 patients with stored electrographic data and appropriate interventions, the interventions were triggered by ventricular tachycardia or fibrillation.

Conclusions: Ventricular tachycardia or fibrillation appears to be the principal mechanism of sudden death in patients with hypertrophic cardiomyopathy. In high-risk patients with hypertrophic cardiomyopathy, implantable defibrillators are highly effective in terminating such arrhythmias, indicating that these devices have a role in the primary and secondary prevention of sudden death.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Cardiomyopathy, Hypertrophic / complications
  • Cardiomyopathy, Hypertrophic / therapy*
  • Child
  • Death, Sudden, Cardiac / etiology
  • Death, Sudden, Cardiac / prevention & control*
  • Defibrillators, Implantable* / adverse effects
  • Equipment Failure
  • Female
  • Humans
  • Male
  • Middle Aged
  • Retrospective Studies
  • Risk Factors
  • Tachycardia, Ventricular / etiology
  • Tachycardia, Ventricular / prevention & control
  • Ventricular Fibrillation / etiology
  • Ventricular Fibrillation / prevention & control