Digital slide and virtual microscopy based routine and telepathology evaluation of routine gastrointestinal biopsy specimens
  1. B Molnar1,
  2. L Berczi2,
  3. C Diczhazy2,
  4. A Tagscherer1,
  5. S V Varga1,
  6. B Szende2,
  7. Z Tulassay1
  1. 1Digital Microscopy Laboratory, Second Department of Medicine, Semmelweis University, Szentkirályi St. 46, 1088 Budapest, Hungary
  2. 2First Department of Pathology and Experimental Oncology, Semmelweis University
  1. Correspondence to:
 Dr B Molnar, II Department of Medicine, Semmelweis University, Szentkirályi St. 46, 1088 Budapest, Hungary;
 mb{at}bel2.sote.hu

Abstract

Aims: To evaluate a recently developed digital slide and virtual microscope system, and to compare this method with optical microscopy on routine gastrointestinal biopsy specimens in both local and remote access modes.

Methods: A fully computer controlled commercial microscope was used. The scanning program included object detection, autofocus, and image compression algorithms. The overall hard disk space for a gastric biopsy was between 30 and 50 MB and the scanning time was between 20 and 40 minutes. Haematoxylin and eosin stained routine gastric (61) and colon (42) biopsy specimens were selected, scanned, and evaluated by two specialists on an optical (OM) and virtual microscope (VM).

Results: The overall concordance of VM and OM with the consensus diagnosis was 95.1% and 97%, respectively. Clinically important concordance was 96.1% and 98% for VM and OM, respectively. The two methods showed concordance in 92% of cases and clinically important concordance in 94.1% of cases. The reasons for discordance were image quality (one case), interpretation difference (three cases), and insufficient clinical information (three cases). Remote evaluation of the digital slides through the Internet has the advantages of the previously used static and dynamic telepathology methods.

Conclusions: Diagnostic concordance was found between OM and VM. The digital slide and the virtual microscope can be alternative techniques in the computerisation of the histology laboratory and in teleconsultation services after further evaluation of time and storage constraints.

  • digital slide
  • digital microscope
  • gastric biopsy
  • section
  • OM, optical microscopy
  • VM, virtual microscopy

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

As a result of the rapid developments in information technology, medical digital image analysis is becoming a widely accepted technique.1 Digital techniques and laboratories2 already exist for radiological applications. New stand alone automated microscope systems have been developed for histological and cytological applications in the past decade.

The automated screening and rescreening of cervical smears is now available for routine practice.3,4 In addition, several new semiautomatic microscopes have been developed to aid in quality control of the cytotechnologists’ work.5–8 These systems notify the observer of the scanned areas and images, and electronic recording of selected images is also available.

Automated histological analysis has been an important research tool for several years, but until recently it was not suitable for routine applications.9,10 However, now histological diagnoses can be supported by new electronic techniques, such as TV image cytometry and teleconsultation based on histological images rather than entire samples.11

“Automated histological analysis has been an important research tool for several years, but until recently it was not suitable for routine applications”

Telepathology services were built up during the past decade around two technology platforms. Dynamic telepathology uses remote controlled microscope systems with high throughput online image transport channels.12,13 This method has the advantage of entire slide access and lacks the error source of preselected microscopical frames, although the costs of setting up and running the system are relatively high. The use of static preselected images for teleconsultation needs much less hardware investment, but sampling errors can occur.14,15 Using the Internet as a telecommunication pathway for static images is a low cost widely available alternative.16,17

The applications of digital slides are considerable, as sources and targets of telepathology and automated histological analysis; however, attempts to prepare electronic or digital slides have been limited (R Ferreira et al. Digital dynamic telepathology—the virtual microscope. Proceedings of the 1997 AMIA Annual Fall Symposium, 1997).18,19 Only recently has the storage capacity and speed of personal computers become sufficient to handle the extremely large amount of image information stored on a slide. In addition, low cost, commercial motorised microscopes have recently been developed by several manufacturers.

Working on digital slides requires virtual or digital microscopy. Preliminary positive results on a limited number of mosaicked microscopic images have been reported recently.20

The aim of our present study was to evaluate our recently reported digital slide and microscopy system21 on routine gastrointestinal biopsy specimens. We compared the performance of optical and digital microscope evaluations in local and remote modes.

MATERIALS AND METHODS

Gastric routine biopsy specimen analysis

Biopsy specimens were placed in buffered formalin, routinely processed, and stained with haematoxylin and eosin. Altogether, 103 specimens were selected from the files of the first department of pathology. Single representative slides were evaluated from each case. Table 1 shows the distribution of the cases.

Table 1

Diagnoses

The histological sections were evaluated in separate settings on a virtual microscope (VM) and an optical microscope (OM), without knowledge of the results of the previous analysis by two independent, board certified histologists.

First, the glass slides were assessed by optical microscopy. The paper copies of the clinical histories and the evaluation report were collected.

Over the next few weeks the slides were digitised; this was carried out in the digital microscopy laboratory. The scanning computer was also used as a digital slide server. The virtual microscopy evaluation was done on a separate local area network workstation in the pathology department with access to the slide server computer using the Internet.

Data analysis

At the end of the study, when all the optical and virtual microscopy results were available, the consensus data, concordance, and source of diagnostic discordance were determined.

Concordance was designated level “A”. Levels “B” and “C” were clinically unimportant and important discordance, respectively, as suggested by Weinberg et al.22

Those cases with discordant results were collected together with the optical and electronic data. The final diagnosis and definition of the source of the discordant data were agreed upon at a consensus session in the consultation room. Here, access to the digital slides was also available through a computer workstation and the Internet.

The reasons for discordance were classified by the consensus meeting as follows: inadequate image quality (class I), interpretation (class II), and insufficient clinical information (class III).

The significance of the concordance was determined using Kendall’s concordance coefficient determined by the Statistica program package (V.4.3, Statsoft, Tulsa, Oklahoma, USA).

Slide digitisation and the virtual microscope system

Hardware tools used

We used an Axioplan 2 MOT (Carl Zeiss, Jena, Germany) microscope. The microscope functions (objectives, stage, focus, illumination, and filters) can be controlled and changed through the RS232 interface from an application program. The mechanical accuracy of the motorised scanning stage for X/Y and Z directions was 0.5 μm.

We used the Grundig FAC 830 1/2 inch, one chip CCD camera with 752 × 581 (440 000) pixels. The integration time of the camera can be controlled through the computer interface RS485. The applied image digitisation card was the Screen Machine II from Fast Electronic Germany (Munich, Germany) with a resolution of 640 × 560 pixels and 64 K colour depth. The programs were running on computers with at least an Intel Pentium II 350 MHz processor, 128 MB RAM, and 2 GB hard disk.

Features of the digital slide scanning program

Scanning area determination was performed by upper right and lower left corner fixing.

Autoscanning was started after setting the stage at the zero position. During the scanning process autofocusing was done only at each third to fifth field of view. All the images were compressed in JPEG format and stored in the slide databank in the corresponding position. Autofocusing was done using Brenner’s algorithm.23 At ×40 magnification, 125 856 frames would need to be stored for the entire slide. However, a threshold filter was used to store only those frames with image content. In this way, only the area containing the biopsy was stored, amounting to 302 to 1334 frames for each slide.

First, the fields of view were mosaicked using mathematical algorithms24; however, the error between the required and real positions of the stage was found to be less than ± 0.5 μm. Based on this high precision no software mosaic alignment was used.

Features of the virtual microscope program

Slide selection

After scanning, the slides were stored in subdirectories called projects for a higher ordering. After selection an electronically minimised slide image is shown on the screen (fig 1).

Figure 1

The user interface of the virtual microscope program. The artificial magnification (×10) of a scanned gastric biopsy section is shown on the screen. Note the moving arrows in the left bottom corner and the + and – labels in the middle. These can be used for discrete step magnification and the reduction of a specified area of the slide. On the right bottom of the screen the scanned area of the slide is highlighted.

Slide orientation map

This map represents the whole slide, where one pixel on the screen corresponds to one field of view. The recorded segments are labelled with white pixels on a grey background.

Applicable electronic magnifications

After finding the area of interest, the user has several options for magnification of the selected segments. The prepared magnification steps of ×100 and ×200 or special “+” and “−” mouse arrows can be used to change the magnification (figs 1–3).

Figure 2

An intermediate magnification of a scanned gastric biopsy specimen (original magnification, ×200).

Figure 3

The real world scanning magnification (original magnification, ×400) on the virtual microscope. The borders between the mosaicked microscopic field of views can also be detected.

Moving and scrolling the slide

If the interested area is not on the screen at the required magnification, then the moving arrows (up, down, left, and right) can be used to move the slide in any direction.

Labelling of interesting frames for re-evaluation, consultation, and reporting

Up to 10 coloured labels can be placed on the digital slide in the software. This option can also be used for reconsultation by experts via the local area network or in specific cases via the Internet.

Internet access

Every virtual microscopy workstation can be a slide server too. On the remote workstation the slide server’s IP address has to be defined. After connecting to the server computer, a list of available slides and the information about these slides is transported to the client workstation. After selecting a slide, its electronically compressed low resolution image is transferred to the workstation. Every image that is transported during the evaluation will be stored on the local machine for safety reasons.

RESULTS

Concordance

Discordance was found in eight cases (7.8%) between the VM or OM results and the consensus diagnosis. In five cases the OM and in three cases the VM yielded the correct diagnosis.

There was concordance in 95 (92.2%) of the 103 cases, and clinically important concordance in 97 cases (94.1%). OM yielded higher (100 of 103; 97%) concordance with the consensus results than did VM (98 of 103; 95.1%). The highest degree of concordance for clinically important diagnosis was also seen for OM (101 of 103; 98%) (table 2).

Table 2

Concordance between optical microscopy, virtual microscopy, and the consensus diagnosis

Reasons for diagnostic discordance

Because the entire specimen was digitised there was no sampling error. However, the other common sources of error in telepathology—low image quality (one of eight), interpretation (four of eight), and insufficient clinical information (three of eight) were seen.

The effects of poor image quality were seen in only one case, although this was a clinically important one. In this case, the VM diagnosis was ulceration ventriculi and the OM diagnosis was ulcerated well differentiated adenocarcinoma ventriculi. The correct diagnosis was achieved by evaluating multiple focus levels in OM, an option that is not currently available in VM (table 3).

Table 3

Reasons for diagnostic discordance

Technical and practical data regarding the application of VM

The hard disk volume of a microscopic field of view is between 60 and 100 KB after JPEG compression. The overall hard disk space required for a gastric biopsy is between 30 and 50 MB and the scanning time is between 20 to 40 minutes, depending on the number and area of sections on a slide.

Evaluating the specimen on the monitor is more comfortable for the histologist, more reproducible, and more easily documented when compared with the optical method (table 4).

Table 4

Comparison between optical and virtual microscope evaluation methods

Remote access via the Internet was relatively fast. The first minimised image was uploaded in real time. As the user switched over to high magnification the single fields of view were seen on the screen in seconds (figs 2,3).

DISCUSSION

In our study, we evaluated the technical feasibility and diagnostic concordance of a digital histological section evaluation system and compared it with the routine OM procedure.

Low cost computational techniques were used. The scanning speed was too slow for routine applications, but there are free resources available for increasing scanning speed by several magnitudes. The digitisation card should be changed to a higher resolution and speed (×10). The computer should also be updated (×5). In addition, an automated microscope could be considered.

Image quality should be further enhanced using a digital one chip, analogue three chip camera instead of the analogue, one chip camera used at present. In this case, an increase in diagnostic accuracy would be expected.

However, these developments and results show that in the near future we can have microscope free virtual microscopy workstations with the functions of the optical microscope. One advantage would be a more comfortable working environment for histologists. In addition, it would also support quality control techniques and consultation with remote experts. The first application of digital slides and virtual microscopy should be telepathology consultation. However, later this technology could be used in the routine histology evaluation process, as was done in our study.

Remote evaluation of the slides through the Internet has the advantages of the previously used static or dynamic telepathology methods (the entire slide is available at high magnification, microscope and remote assistance free evaluation).25

We found a higher concordance with optical microscopy than previous static image based telepathology studies.13–15 This might be explained by the fact that this technology eliminates sampling error.

Time constraints and fatigue were not evaluated. We are planning to assess these factors in a larger interlaboratory study.

“In the near future we can have microscope free virtual microscopy workstations with the functions of the optical microscope”

The reasons for discordance yielded new information for future developments. In selected cases several focus planes are required for correct diagnosis. The virtual microscope cannot be used for the analysis of underlying cells. However, with further technical and mathematical developments several focus levels can be recorded, stored, and evaluated by the virtual microscope.

In his 1996 review article, O’Brien stated that the computerisation of the histology laboratory was desirable but that it would be far in the future.26 More, recently, Leong and McGee27 stated that complete automated slide digitisation has influence at all levels of clinical practice and education. They emphasised the importance of dedicated software technology. Recently, the usefulness of digital slides and virtual microscopy for quality assurance and teaching applications has been demonstrated.28,29

Our results show that digital slides and virtual microscopy technology can be used in selected cases for telepathology consultation. After the definition of the time and storage requirements the routine use of this technique could also be considered.

Take home messages

  • We found diagnostic concordance between optical microscopy and virtual microscopy on gastrointestinal biopsy specimens

  • Virtual microscopy has the advantage of providing a more comfortable working environment for histologists

  • Virtual microscopy also supports quality control techniques and consultation with remote experts

  • Initially, digital slides and virtual microscopy should be used in telepathology consultation

  • After further evaluation of time and storage constraints this technology might be useful in routine histology